Product Information
Technical specifications:
- Exterior Dimensions : 4' x 2' x 2'
- Work Surface : SS 304 , ( Removable & chemical resistant ).
- Exterior Body : Stainless Steel
- Side and Walls : Stainless Steel
- Front Sash Type : Vertical sliding, 10° angled sash
- Working Opening Height : 10"
- Sash Material : Shatterproof (¼" ) Made of toughened safety glass
- Filtration Type : HEPA filters for supply ( down flow) and exhaust air
- HEPA Filter Efficiency : ≥ 99.99% at 0.3 microns
- Control System : Fully Microprocessor-based automatic control & supervision.
- Airflow Type : Vertical laminar downflow
- Average Inflow Velocity : ≥100 FPM
- Downflow Velocity : ≥ 55 FPM
- Air Recirculation Ratio : 70% Re-circulated and 30% exhausted
- Display : Digital display for airflow parameters, system status with alarms
- Alarm System : Audio-visual alarms for airflow failure, sash position, filter clogging, and system faults
- Electrical Outlet : Internal electrical socket compatible with Indian plug-top
- Power Supply : 220–240 V AC, 50–60 Hz, single phase
- Noise Level : Low noise operation suitable for lab. environment
- Safety Features : Airflow interlocks, sash interlocks, alarms, and fail-safe controls
- Warranty : 3 year’s
BIO SAFETY CABINET (CLASS A2)
Salient Features:
- Class II, Type A2 Biological Safety Cabinet.
- As per NSF 49, EN 12469 standards.
- Motor & fan located on the clean side of the cabinet.
- Differential pressure indicator for HEPA filter condition.
- Side & rear walls Made of Single Piece SS 304 with coved corners.
- Heavy-duty telescopic leg assembly with height adjustment Stand
- DOP test ports for HEPA filter leak testing.
- Plenum & interior zones maintained under negative pressure.
- Shatterproof sash with unobstructed viewing area.
- Suitable agent ( Formaldehyde / Hydrogen Peroxide – VHP).
- Dedicated port provided for cabinet decontamination .
- Automatic compensation for voltage fluctuations to maintain & blower with motor speed against filter loading.
- Automatic blower idling when sash is fully closed while maintaining airflow conditions.
- Unobstructed, full-width clear viewing Area.
- Floor-mounted with telescopic stand.
- Ergonomic design for operator comfort & reduced fatigue.
- Audio-visual alarm when sash exceeds safe working height.