Product Information
Technical Parameters:
Capacity
120 liters
Temp. Range
Ambient 5°C to 50°C with 0.1°C resolution
Inner Chamber Material
High-grade Stainless steel 304 Grade (smooth, easy-to-clean)
Temp. Accuracy
±0.1°C
CO? Range
0-20%
CO? Control Accuracy
±0.1%
Relative Humidity
90-95% RH (Maintains)
Gas Mixing System
Integrated CO? gas mixing jet with Venturi-effect technology.
Access Ports
Two Nos. ports (Ø ?25 mm) Side for external probes & Sensors
Display
Large backlit display temp. monitoring, control, & alarm indication.
Chamber Uniformity
±0.4°C at 37°C
Heating System
Six-Sided Direct heating for uniform temp. distribution
Temperature Recovery
37°C within ?6 min. (30-Sec.) Door opening
CO? Recovery
5% CO? within ?5 min. after 30-sec. Door Opening
Humidity Recovery
Rapid Humidity Recovery after door opening.
Temperature Sensor
Independent PT100 temp. sensor.
Co? Sensor
Dual-channel NDIR (Infrared) CO? sensor (Automatic calibration)
Sensor Durability
CO? Sensor Suitable for ?140°C High-temp. sterilization.
Controller Accuracy
0.1°C Setting & display resolution.
Gas Regulation
Inline pressure regulator reduced CO? consumption & Stable Pressure Control
Display
High-resolution LCD display.
Connectivity
Ethernet and USB port for data download.
Temp. Display
Temp. indication in °C.
Set Point Wait Function
Process timer starts only after the set temp. is achieved
Dry Heat Sterilization
Up to 125°C ?4 hrs Max. cycle time.
Display & Alarms
Digital display for temp. & alarm monitoring.
Power Supply
230V ±10%, AC 50Hz
Certification
ISO 9001, CE Certified
Documentation
IQ, OQ and PQ Documentation and NABL Accredited Lab Approved Certified
Includes Accessories
1 Nos. CO? Cylinder, 2 Nos. Pressure Regulators. and Instructional Manual.
Co2 INCUBATOR (PSAW-120)
Description:
- PSAW Co2 Incubator is suitable for cell culture applications, maintaining optimal temp., humidity, & Co2 concentration to ensure ideal pH and sterile conditions.
Salient Features:
- Precisely maintains 37°C temp., high humidity, and controlled CO? levels.
- In-situ 140°C High-temp. Sterilization without removal ( heat-up, sterilization, cool-down to 37°C) 14 hrs.
- Large backlit display temp. monitoring, control, & alarm indication.
- Single inner tempered glass door for sample observation without disturbing chamber conditions.
- Air-jacketed Heating with microprocessor control.
- Seamless, deep-drawn SS inner chamber; no welds or corners.
- Removable tray for easy cleaning and refilling.
- Fan-less operation for low contamination and noise.
- Corrosion-resistant system Construction and components
- Automatic over-temperature alarm with visual indication.
- Adjustable electronic over-temp. safety controller.
- Temp. safety device compliant with DIN 12880.