Product Information
Technical Specifications:
- MOC - Double-walled insulated construction.
- Inner chamber- SS 304
- Outer body - MS powder-coated (corrosion-resistant material)
- Lid Type - Hinged insulated lid with handle
- Plateform & Grip - SS Heavy-duty perforated
- Bottom Support - SS perforated tray
- Mobility - Mounted on sturdy legs
- Applicable Standard - IS 3495 (Part 2)
- Temperature Range - Ambient to 50°C
- Temperature Resolution - 0.1°C
- Temperature Accuracy - ±1°C
- Temperature Uniformity - ±2°C
- Temperature Sensor - PT-100
- Temperature Controller - Microprocessor-based Digital PID Controller
- Heating System - Electric immersion heaters
- Working Chamber Size - 190 × 90 × 90 mm
- Drainage - Bottom drain valve
- Insulation - PUF
- Power Supply - 230 V AC, Single Phase, 50 Hz
- Warranty - 1 Year
Safety Features:
- Over-temperature protection.
- Dry-run protection for heaters.
- Electrical overload and short-circuit protection.
- Proper earthing arrangement.
- Insulated external surfaces to minimise heat loss and operator hazards.
Standard Accessories
- SS perforated specimen platform.
- Water circulation system.
- Drain valve and water inlet fittings.
- Power cable with plug.
- User manual.
- Calibration certificate for temperature controller.
WATER BATH FOR WATER ABSORPTION TEST
Salient Features:
- Suitable for conducting water absorption tests on bricks in accordance with IS 3495 (Part 2) standard.
- Robust, corrosion-resistant construction and designed for continuous laboratory operation.
- Proper drainage arrangement with drain valve for easy cleaning.
- Water inlet provision for convenient filling.
- Circulation pump for uniform temperature distribution
- Suitable for accommodating minimum 24 standard brick specimens.
- Maintain uniform water temperature throughout the chamber.
- Accommodate standard brick specimens without overcrowding.
- Permit easy loading and unloading of specimens.
- Ensure continuous operation without significant temperature fluctuation.
- Suitable for prolonged immersion testing as per IS requirements.